Recalled by RTI Surgical, Inc. (Dba Pioneer Surgical Technology, Inc.) — Marquette, MI

FDA device recall Z-0290-2017 was initiated by RTI Surgical, Inc. (Dba Pioneer Surgical Technology, Inc.) on Sep 16, 2016 and is designated Class II. Reason for recall: RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on… The recall status is terminated (terminated Jun 1, 2017). Affected quantity: 66.

Recall details

Product Type
Devices
Report Date
Oct 26, 2016
Initiation Date
Sep 16, 2016
Termination Date
Jun 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
66

Product description

RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Usage: When intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (occiput-T3), the Streamline OCT Occipito-Cervico-Thoracic System is intended for: degenerative disc disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/ dislocation, atlanto/axial fracture with instability, occipito-cervical dislocation, deformities or curvature, tumors, pseudoarthrosis, and revision of previous cervical and upper thoracic spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the pedicle screws (standard and high angle) is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The pedicle screws are not intended for use in the cervical spine. The hooks, connectors, and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.

Reason for recall

RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants could cause reduced strength of these affected implants, and lead to patient injury.

Distribution pattern

US Nationwide Distribution in the states of: AL, CA, CO, FL, IN, IL, MI, MO, MS, NC, NY, OK, TN, TX, and VA,.

Code information

Part Number: 23-CC-ROD-240; Lot Number: 234829, UDI Number: (01)0084646802 7179( 10}234829; Lot Number: 246678, UDI Number: {01}00846468027179{10}246678

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0290-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for RTI Surgical, Inc. (Dba Pioneer Surgical Technology, Inc.)? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for RTI Surgical, Inc. (Dba Pioneer Surgical Technology, Inc.): recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about RTI Surgical, Inc. (Dba Pioneer Surgical… →

Source: openFDA record for Z-0290-2017. Report a data issue: hello@regdatalab.com.