Z-0291-2025 Class I Ongoing
Recalled by Philips Respironics, Inc.
— Murrysville, PA
FDA device recall Z-0291-2025 was initiated by Philips Respironics, Inc. on October 7, 2024 and is designated Class I. Reason for recall: Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements. The recall status is ongoing. Affected quantity: 82,800 units.
Recall Details
- Product Type
- Devices
- Report Date
- November 20, 2024
- Initiation Date
- October 7, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 82,800 units
Product Description
Trilogy Evo. Model Numbers: DS2110X11B, UDS2110X11B, RDS2110X11B, CA2110X12B, JP2110X16B, IN2110X15B, DE2110X13B, AU2110X15B, BL2110X15B, EU2110X15B, BR2110X18B, LA2110X15B, EE2110X15B, FR2110X14B, GB2110X15B, ND2110X15B, ES2110X15B, IA2110X15B, IT2110X21B, KR2110X15B, TR2110X15B, EU2110X19, FP2110X10, PP2110X10, RBRBR2110X18B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Reason for Recall
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of AE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, LY, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, TZ, UA, UY, VN, YT, ZA, ZW.
Code Information
Model Number (UDI-DI Number): AU2110X15B (606959051980); BL2110X15B (606959055520); BR2110X18B (606959055476); CA2110X12B (606959052307); DE2110X13B (606959054110); DS2110X11B (606959051942); EE2110X15B (606959055544); ES2110X15B (606959054080); EU2110X15B (606959054981); EU2110X19 (606959055537); FP2110X10 (606959062092); FR2110X14B (606959055506); GB2110X15B (606959054127); IA2110X15B (606959055490); IN2110X15B (606959051959); IT2110X21B (606959055421); JP2110X16B (606959051973); KR2110X15B (606959055483); LA2110X15B (606959055452); ND2110X15B (606959055513); PP2110X10 (606959054035); RBRBR2110X18B (606959055476); RDS2110X11B (606959060517); TR2110X15B (606959055551); UDS2110X11B (606959060500). All Serial Numbers.