Z-0293-2017 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-0293-2017 was initiated by Philips Electronics North America Corporation on Sep 19, 2016 and is designated Class II. Reason for recall: Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX The recall status is terminated (terminated May 25, 2018). Affected quantity: 2,212 units.

Recall details

Product Type
Devices
Report Date
Oct 26, 2016
Initiation Date
Sep 19, 2016
Termination Date
May 25, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,212 units

Product description

Philips IntelliVue MX40 Patient Monitor: IntelliVue MX40 802.11a/b/g Product Number: 865352 Exchange part: 453564262491 453564262511 453564615311 453564615331 453564262571 453564262591 Exchange part: 453564262531 453564262551 453564262571 453564262591 453564615351 453564615371 453564467761 (Korea) 453564467781 (Korea) 453564451791 (Japan, Taiwan, Indonesia) 453564451811 (Japan, Taiwan, Indonesia) 453564467821 (Mexico) 453564467841 (China) 453564467861 (China) Worldwide SW Revisions: B.05.28, B.05.29, and B.05.32 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals

Reason for recall

Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iX

Distribution pattern

Worldwide Distribution-US (Nationwide) and countries of: Aruba, Australia, Austria, Belgium, Bulgaria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iraq, Ireland, Italy, Japan, Korea, Republic of, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Singapore, Slovenia, South Africa, Spain Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and United Kingdom.

Code information

SW Revisions: B.05.28, B.05.29, and B.05.32

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0293-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0293-2017. Report a data issue: hello@regdatalab.com.