Z-0294-2024 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-0294-2024 was initiated by Philips North America Llc on Oct 9, 2023 and is designated Class II. Reason for recall: For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS)… The recall status is ongoing. Affected quantity: 29 units.

Recall details

Product Type
Devices
Report Date
Nov 22, 2023
Initiation Date
Oct 9, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29 units

Product description

Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator designed for radiation oncology and therapy.

Reason for recall

For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.

Distribution pattern

Nationwide Foreign: Austria Belgium Canada China Estonia France Germany Hondorus India Italy Mexico Netherlands Philippines Poland Singapore Spain Switzerland Taiwan

Code information

UDI-DI: (01)00884838059450 Serial Numbers: Serial Number 7362 75114 7441 7449 7996 7104 7333 7991 7497 7762 7254 7373 7674 7783 7971 7802 7385 7840 7353 750010 75107 8007 7990 7716 7689 7618 75045 7442 750017

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0294-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America Llc →

Source: openFDA record for Z-0294-2024. Report a data issue: hello@regdatalab.com.