FDA device recall Z-0298-2013 was initiated by Biomet 3I, LLC on Sep 28, 2012 and is designated Class II. Reason for recall: Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too… The recall status is terminated (terminated Aug 6, 2015). Affected quantity: 76 devices.
Recall Details
- Product Type
- Devices
- Report Date
- Nov 28, 2012
- Initiation Date
- Sep 28, 2012
- Termination Date
- Aug 6, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 76 devices
Product Description
Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.
Reason for Recall
Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant.
Distribution Pattern
Worldwide Distribution-USA (nationwide) including the states of VA, SC, HI, LA, OH, FL, IL, NJ, NC, CA, CT, MI, TN, AK, NH, NY, MA, GA, PA, and KY and the countries of United Kingdom, Spain, Germany, Canada, Japan, Singapore, Taiwan, Israel, Poland, Hong Kong, Ireland, The Netherlands, Austria, and France.
Code Information
Model Number IFOA411, Lot # 2011111333, 2011111333-S1