Z-0303-2017 Class II Terminated

Recalled by Handicare USA, Inc. — Allentown, PA

FDA device recall Z-0303-2017 was initiated by Handicare USA, Inc. on Apr 15, 2014 and is designated Class II. Reason for recall: During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the… The recall status is terminated (terminated Dec 13, 2016). Affected quantity: 29 units , plus 47 wing handles were shipped as spare parts.

Recall details

Product Type
Devices
Report Date
Nov 2, 2016
Initiation Date
Apr 15, 2014
Termination Date
Dec 13, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
29 units , plus 47 wing handles were shipped as spare parts

Product description

ReTurn 7500/7500 Aid, transfer

Reason for recall

During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.

Distribution pattern

US Distribution to the states of : NY, NJ, NV, NE, IL, CT, PA and ME

Code information

Model # 7400, 7500 and 7322 (spare part) Serial # 373317697270501334, 373317697270501430, 073317697400500012, 073317697400500057

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0303-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Handicare USA, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0303-2017. Report a data issue: hello@regdatalab.com.