Z-0310-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- November 11, 2020
- Initiation Date
- September 17, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8,872 devices
Product Description
A Li-Gator Single Use Laparoscopic Clip Applier, 10 mm X 29 cm - Product Usage: intended for use on tubular structures or vessel wherever a metal litigating clip is intended
Reason for Recall
Device malfunction causing the applier to jam, and clip loading failures.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of FL, TN, MO, CA, GA, and IN. The countries of Chile, Croatia, Dubai (UAE), Equador, Egypt, France, Germany, Greece, Guatemala, Hong Kong, Iraq, Ireland, Israel, Jordan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Palestinian, Quatar, Saudi Arabia, Slovenia, South Africa, Switzerland, Tunisia, United Kingdom, and Venezuela.
Code Information
Model # 360-010-001 (A Li-Gator) lot codes: I8250 I8251 I8252 I8820 I8821 I8822 I8823 I8824 I8825 I9571-A I9572-A I9573-A I9716-A I9717-A I9718-A I9719-A J0113-A J0114-A J0115-A J0225-A J0227-A J0228-A J0229-A J0230-A J0231-A J0240-A J0243-A J0287-A