Z-0311-2018 Class II Terminated
FDA device recall Z-0311-2018 was initiated by Philips Electronics North America Corporation on Jul 14, 2017 and is designated Class II. Reason for recall: If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of… The recall status is terminated (terminated Jun 5, 2020). Affected quantity: 412.
Recall details
- Product Type
- Devices
- Report Date
- Jan 10, 2018
- Initiation Date
- Jul 14, 2017
- Termination Date
- Jun 5, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 412
Product description
Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.
Reason for recall
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
Distribution pattern
412 units affected.
Code information
9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017.