Z-0311-2018 Class II Terminated

Recalled by Philips Electronics North America Corporation — Andover, MA

FDA device recall Z-0311-2018 was initiated by Philips Electronics North America Corporation on Jul 14, 2017 and is designated Class II. Reason for recall: If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of… The recall status is terminated (terminated Jun 5, 2020). Affected quantity: 412.

Recall details

Product Type
Devices
Report Date
Jan 10, 2018
Initiation Date
Jul 14, 2017
Termination Date
Jun 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
412

Product description

Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.

Reason for recall

If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.

Distribution pattern

412 units affected.

Code information

9890-010-88152 - Codes included in attached list. Devices were manufactured and distributed from March 2015 - March 2017.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0311-2018
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0311-2018. Report a data issue: hello@regdatalab.com.