Z-0314-2027 Class II Ongoing

Recalled by Stryker Leibinger GmbH & Co. KG — Freiburg Im Breisgau

FDA device recall Z-0314-2027 was initiated by Stryker Leibinger GmbH & Co. KG on Sep 24, 2026 and is designated Class II. Reason for recall: Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient. The recall status is ongoing. Affected quantity: 513.

Recall details

Product Type
Devices
Report Date
Nov 4, 2026
Initiation Date
Sep 24, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
513

Product description

Precision Pointer Electromagnetic (8000-050-001)

Reason for recall

Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.

Distribution pattern

Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.

Code information

Lot 7000413361 UDI-DI 07613327413205

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0314-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0314-2027. Report a data issue: hello@regdatalab.com.