Z-0317-2018 Class II Terminated

Recalled by Repro-Med Systems, Inc. — Chester, NY

FDA device recall Z-0317-2018 was initiated by Repro-Med Systems, Inc. on Mar 10, 2016 and is designated Class II. Reason for recall: On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the… The recall status is terminated (terminated Feb 26, 2018).

Recall details

Product Type
Devices
Report Date
Jan 17, 2018
Initiation Date
Mar 10, 2016
Termination Date
Feb 26, 2018
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Freedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275, F600, F45, F1200 F180.

Reason for recall

On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out of 3,195 were rejected due to visual defect, gap in the supplier s manufacturing seal for Part number #317036 5 X 6.5 ST Bag Combo. the firm decided to recall Precision Flow Rate Tubing and RMS HIgH-Flo Subcutaneous Safety Needles Sets due to the sterile barrier packaging being found to have gaps during post-sterilization inspection of one lot of flow tubing, in which 701 out of 3,195 units were rejected for not passing inspection. The firm determined that the issue was large than just the one lot and recalled the product via March 10, 2016 Voluntary Medical Device Corrections and Removal Notification to customers. The recall was conducted without FDA notification.

Distribution pattern

Nationwide in US

Code information

Lot Numbers: 1.053/16, 1.052/16, 1.051/16, 1.050/16, 1.049/16, 1.048/16, 1.047/16, 1.042/16, 1.041/16, 1.040/16, 1.039/16, 1.038/16, 1.037/16, 1.036/16, 1.033/16, & 1.030/16 Expiration: 2021-02

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0317-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Repro-Med Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0317-2018. Report a data issue: hello@regdatalab.com.