Z-0317-2027 Class II Ongoing

Recalled by Accriva Diagnostics, Inc. — San Diego, CA

FDA device recall Z-0317-2027 was initiated by Accriva Diagnostics, Inc. on Sep 14, 2026 and is designated Class II. Reason for recall: Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high… The recall status is ongoing. Affected quantity: 28710.

Recall details

Product Type
Devices
Report Date
Nov 11, 2026
Initiation Date
Sep 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28710

Product description

Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103

Reason for recall

Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.

Distribution pattern

Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.

Code information

UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0317-2027
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Accriva Diagnostics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0317-2027. Report a data issue: hello@regdatalab.com.