Z-0319-2026 Class II Terminated

Recalled by Surgical Theater Inc — Beachwood, OH

FDA device recall Z-0319-2026 was initiated by Surgical Theater Inc on Sep 16, 2025 and is designated Class II. Reason for recall: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image… The recall status is terminated (terminated Aug 3, 2026). Affected quantity: 45 units.

Recall details

Product Type
Devices
Report Date
Nov 5, 2025
Initiation Date
Sep 16, 2025
Termination Date
Aug 3, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
45 units

Product description

SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system

Reason for recall

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Distribution pattern

Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.

Code information

UDI/DI 00863157000336

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0319-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0319-2026. Report a data issue: hello@regdatalab.com.