Z-0328-2027 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

FDA device recall Z-0328-2027 was initiated by GE Medical Systems, LLC on Aug 28, 2026 and is designated Class II. Reason for recall: GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Nov 11, 2026
Initiation Date
Aug 28, 2026
Voluntary/Mandated
FDA Mandated

Product description

Senographe Pristina 3D

Reason for recall

GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).

Distribution pattern

USA and OUS.