Z-0328-2027 Class II Ongoing
FDA device recall Z-0328-2027 was initiated by GE Medical Systems, LLC on Aug 28, 2026 and is designated Class II. Reason for recall: GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter… The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Nov 11, 2026
- Initiation Date
- Aug 28, 2026
- Voluntary/Mandated
- FDA Mandated
Product description
Senographe Pristina 3D
Reason for recall
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Distribution pattern
USA and OUS.
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0328-2027