Z-0332-2022 Class II Ongoing

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
December 15, 2021
Initiation Date
November 12, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units OUS

Product Description

Covidien Sonicision Battery Charger-intended to charge the Sonicision Reusable Battery Pack and to provide information on the battery charging progress and battery pack status Item Code: CBCA

Reason for Recall

Intermittent risk of electric shock to users if an affected charger is used in conjunction with a faulty or bypassed institutions Mains Protective Earth (Ground) circuit

Distribution Pattern

Brazil

Code Information

Serial Numbers: MCSB006939, MCSB006947 GTIN: 10884521592896