Z-0334-2023 Class I Terminated

Recalled by Remel Inc — Lenexa, KS

FDA device recall Z-0334-2023 was initiated by Remel Inc on Oct 20, 2022 and is designated Class I. Reason for recall: An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp… The recall status is terminated (terminated Sep 30, 2024). Affected quantity: 2656 units.

Recall Details

Product Type
Devices
Report Date
Dec 14, 2022
Initiation Date
Oct 20, 2022
Termination Date
Sep 30, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2656 units

Product Description

Thermo SCIENTIFIC, Sensititre STP6F, Gram Negative IVD AST

Reason for Recall

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Distribution Pattern

US Nationwide Distribution

Code Information

Lot Numbers: B2314, B2174, B2036, B1346, B1183, B1115, B1027