Z-0336-2016 Class II Terminated

Recalled by NxStage Medical, Inc. — Lawrence, MA

Recall Details

Product Type
Devices
Report Date
December 2, 2015
Initiation Date
October 29, 2015
Termination Date
November 19, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
133 units

Product Description

NxStage System One S Cycler - Model no. NX1000-3-A (OUS). For home hemodialysis.

Reason for Recall

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Distribution Pattern

Worldwide Distribution.

Code Information

Software version 4.9 and 4.10