Z-0336-2022 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 22, 2021
- Initiation Date
- November 15, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5,566 (US); 6,354 (OUS)
Product Description
Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.
Reason for Recall
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
Distribution Pattern
Worldwide distribution - US Nationwide.
Code Information
All Serial Numbers. May also be designated as part number D998 etc. Model Number, UDI-DI: 0998-00-0800-31, 10607567109053; 0998-UC-0800-31, 10607567109053; 0998-00-0800-32, 10607567111117; 0998-UC-0800-32, 10607567111117; 0998-00-0800-33, 10607567109008; 0998-UC-0800-33, 10607567109008; 0998-00-0800-34, 10607567111940; 0998-UC-0800-34, 10607567111940; 0998-00-0800-35, 10607567109107; 0998-UC-0800-35, 10607567109107; 0998-00-0800-36, 10607567114187; 0998-UC-0800-36, 10607567114187; 0998-00-0800-45, 10607567108421; 0998-UC-0800-45, 10607567108421; 0998-00-0800-52, 10607567108438; 0998-UC-0800-52, 10607567108438; 0998-00-0800-53, 10607567108391; 0998-UC-0800-53, 10607567108391; 0998-00-0800-55, 10607567108414; 0998-UC-0800-55, 10607567108414; 0998-00-0800-65, 10607567113432; 0998-UC-0800-65, 10607567113432