Z-0342-2022 Class I Terminated

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
December 15, 2021
Initiation Date
November 4, 2021
Termination Date
July 9, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
278 total

Product Description

Puritan Bennett 980 Series Ventilator, 980U1ENASAA

Reason for Recall

Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

Distribution Pattern

Worldwide distribution - US Nationwide.

Code Information

GTIN/UDI - 10884521201385 Serial Numbers: 35B1700252 35B1700365 35B1700399 35B1700407 35B1700489 35B1700505 35B1700513 35B1700555 35B1700557 35B1700562 35B1700563 35B1700565 35B1700566 35B1700605 35B1700612 35B1700634 35B1700651 35B1700662 35B1700663