Z-0343-2018 Class I Terminated

Recalled by Vyaire Medical — Vernon Hills, IL

FDA device recall Z-0343-2018 was initiated by Vyaire Medical on Nov 29, 2014 and is designated Class I. Reason for recall: Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water… The recall status is terminated (terminated Feb 5, 2021). Affected quantity: 16,670 units in total.

Recall Details

Product Type
Devices
Report Date
Jan 31, 2018
Initiation Date
Nov 29, 2014
Termination Date
Feb 5, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16,670 units in total

Product Description

Infant AirLife single-limb, continuous-flow circuit (Model AH132) containing AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

Reason for Recall

Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

Distribution Pattern

US Nationwide Distribution

Code Information

CM17H12 CM17H16 CM17H17 CM17H19 CM17H20 CM17J27 CM17J28 CM17J29 CM17J30 CM17K09 CM17K20 CM17K21 CM17K22 CM17K23

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0343-2018
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.