Z-0345-2016 Class II Terminated

Recalled by DeRoyal Industries Inc — Powell, TN

FDA device recall Z-0345-2016 was initiated by DeRoyal Industries Inc on Oct 14, 2015 and is designated Class II. Reason for recall: DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper… The recall status is terminated (terminated Jul 1, 2017). Affected quantity: 16 cases of 50 units (800 eaches).

Recall details

Product Type
Devices
Report Date
Dec 9, 2015
Initiation Date
Oct 14, 2015
Termination Date
Jul 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 cases of 50 units (800 eaches)

Product description

HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.

Reason for recall

DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.

Distribution pattern

Distributed in the states of VA, AL, FL, and CA.

Code information

Lot Numbers: 39098746, 39655101, 39985370, 40067795

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0345-2016
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for DeRoyal Industries Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for DeRoyal Industries Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about DeRoyal Industries Inc →

Source: openFDA record for Z-0345-2016. Report a data issue: hello@regdatalab.com.