Z-0345-2020 Class II Terminated

Recalled by Abiomed, Inc. — Danvers, MA

FDA device recall Z-0345-2020 was initiated by Abiomed, Inc. on Sep 4, 2019 and is designated Class II. Reason for recall: This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10… The recall status is terminated (terminated Jan 21, 2020). Affected quantity: 1.

Recall details

Product Type
Devices
Report Date
Nov 13, 2019
Initiation Date
Sep 4, 2019
Termination Date
Jan 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product description

lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI

Reason for recall

This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.

Distribution pattern

The product was distributed to GA

Code information

UDI: (01)00813502011371(10)1430951(17)210331(21)192148A; Serial Number: 192148A

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0345-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abiomed, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0345-2020. Report a data issue: hello@regdatalab.com.