Z-0345-2020 Class II Terminated

Recalled by Abiomed, Inc. — Danvers, MA

Recall Details

Product Type
Devices
Report Date
November 13, 2019
Initiation Date
September 4, 2019
Termination Date
January 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI

Reason for Recall

This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.

Distribution Pattern

The product was distributed to GA

Code Information

UDI: (01)00813502011371(10)1430951(17)210331(21)192148A; Serial Number: 192148A