Z-0362-2020 Class II Terminated

Recalled by Tec Com Gmbh — Kabelsketal, N/A

Recall Details

Product Type
Devices
Report Date
November 20, 2019
Initiation Date
August 14, 2018
Termination Date
December 20, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product Description

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.

Reason for Recall

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

Distribution Pattern

US Nationwide distribution in the state of UT.

Code Information

Serial Numbers: 2018/09/07, 2018/10/01, 2018/10/02, 2018/10/03, 2018/10/05