Z-0363-2022 Class II Ongoing

Recalled by Smith & Nephew, Inc. — Memphis, TN

Recall Details

Product Type
Devices
Report Date
December 22, 2021
Initiation Date
November 1, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38 units

Product Description

smith&Nephew EVOS MEDIAL DISTAL FEMUR PLATE, L, 5H, 90MM, REF 72573101

Reason for Recall

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

Distribution Pattern

US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.

Code Information

Batch Number: 20HM10249, 20HM10250, 20HM10251, 20HM22783, UDI (01) 00885556752999 (17) 310630 (10) 12345678 (11) 2210630 (21) 87654321