Z-0365-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 14, 2022
- Initiation Date
- August 31, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 38 units
Product Description
FULL VISION INC. TRACKMASTER,TREADMILL TMX428 220V, Model #317-07927
Reason for Recall
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Distribution Pattern
Wi, GA, NY, Mexico
Code Information
UDI/DI 00860176000613, Serial Numbers: FVDC-7630, FVDC-7631, FVDC-7633, FVDC-7634, FVDC-7635, FVDC-7636, FVDC-7637, FVDC-7638, FVDC-7639, FVDC-7647, FVDC-7648, FVDC-7649, FVDC-7650, FVDC-7651, FVDC-7652, FVDC-7653, FVDC-7654, FVDC-7655, FVDC-7656, FVDC-7657, FVDC-7658, FVDC-7659, FVDC-7660, FVDC-7661, FVDC-7662, FVDC-7663, FVDC-7664, FVDC-7665, FVDC-7707, FVDC-7708, FVDC-7709, FVDC-7710, FVDC-7711, FVDC-7712, FVDC-7713, FVDC-7714, FVDC-7715, FVDC-7716