Z-0366-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 14, 2022
- Initiation Date
- August 31, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product Description
FULL VISION INC. TRACKMASTER, TREADMILL TMX428CP 110V, Model #317-07928
Reason for Recall
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Distribution Pattern
Wi, GA, NY, Mexico
Code Information
UDI/DI 00860176000620, Serial Numbers: FVDC-7644, FVDC-7645, FVDC-7646, FVDC-7686