Z-0366-2027 Class II Ongoing

Recalled by Altivera Medical Inc. — Englewood, CO

FDA device recall Z-0366-2027 was initiated by Altivera Medical Inc. on Jun 1, 2025 and is designated Class II. Reason for recall: Muscle stimulator may deliver constant electrical impulses to all four electrodes simultaneously, instead of alternating between the electrodes as designed, which may cause muscle fatigue. The recall status is ongoing. Affected quantity: 1238.

Recall details

Product Type
Devices
Report Date
Nov 18, 2026
Initiation Date
Jun 1, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1238

Product description

Zynex Medical M-Wave, Model: 701200

Reason for recall

Muscle stimulator may deliver constant electrical impulses to all four electrodes simultaneously, instead of alternating between the electrodes as designed, which may cause muscle fatigue.

Distribution pattern

US Nationwide distribution including in the states of TX, TN, NJ, OH, IA, SC, WI, GA, MD, FL, PA, CO, NY, OK, CT, AR, KY, KS, CA, AL, DE, LA, VA, NM, NE, MN, IN, MI, MO, DC, WA, UT, IL, ME, RI, MA, WV, NC, NV, MS, HI, OR, ND, AK, AZ, NH, ID, VT.

Code information

UDI-DI: 00852670007540. Firmware Version: 4.3

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0366-2027
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.