Z-0370-2023 Class II Ongoing

Recalled by Full Vision Inc — Newton, KS

FDA device recall Z-0370-2023 was initiated by Full Vision Inc on Aug 31, 2022 and is designated Class II. Reason for recall: This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether… The recall status is ongoing. Affected quantity: 8 units.

Recall details

Product Type
Devices
Report Date
Dec 14, 2022
Initiation Date
Aug 31, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8 units

Product description

FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Reason for recall

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Distribution pattern

Wi, GA, NY, Mexico

Code information

UDI/DI 00860176000675, Serial Numbers: GEDC-6765, GEDC-6766, GEDC-6767, GEDC-6768, GEDC-6849, GEDC-6850, GEDC-6851, GEDC-6852

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0370-2023
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Full Vision Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Full Vision Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Full Vision Inc →

Source: openFDA record for Z-0370-2023. Report a data issue: hello@regdatalab.com.