Z-0372-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 16, 2016
- Initiation Date
- September 27, 2016
- Termination Date
- July 17, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10 replacement cylinders.
Product Description
Replacement floor locking cylinders for the Schaerer model 7300 surgical tables. Replacement cylinders are part number 85118. Product Usage: The cylinders are used to lock the surgical table in place.
Reason for Recall
The floor locking cylinders are contained in the base of the Schaerer 7300 model surgical tables. The cylinders were manufactured with the incorrect "O" ring which may cause the cylinder to leak.
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of: CA, CT, FL, IA, IN, MI, MN, MO, OH, PA, and VA Replcement cylinders wre distributed
Code Information
Lot code of the replacement cylinders, 2016-02-01