Z-0372-2026 Class II Ongoing
FDA device recall Z-0372-2026 was initiated by ASP Global, LLC. dba Anatomy Supply Partners, LLC. on May 5, 2025 and is designated Class II. Reason for recall: All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Sy… The recall status is ongoing. Affected quantity: 4,272,858.
Recall Details
- Product Type
- Devices
- Report Date
- November 5, 2025
- Initiation Date
- May 5, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,272,858
Product Description
SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 ¿L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 ¿L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL¿ End-to-End Capillary Tubes, 30 ¿L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910
Reason for Recall
All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.
Code Information
Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.