Z-0378-2016 Class II Terminated

Recalled by LeMaitre Vascular, Inc. — Burlington, MA

FDA device recall Z-0378-2016 was initiated by LeMaitre Vascular, Inc. on Oct 21, 2015 and is designated Class II. Reason for recall: Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility. The recall status is terminated (terminated Jul 12, 2019). Affected quantity: 4315 total.

Recall details

Product Type
Devices
Report Date
Dec 16, 2015
Initiation Date
Oct 21, 2015
Termination Date
Jul 12, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4315 total

Product description

LeMaitre Over-the-Wire Embolectomy Catheter 3F, 40cm, Model 1651-34 and e1651-34. Indicated for the removal of arterial emboli and thrombi.

Reason for recall

Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.

Distribution pattern

Worldwide Distribution. US nationwide including PUERTO RICO, and Australia, AUSTRIA, BELGIUM, COLOMBIA, COSTA RICA, Czech Republic, DENMARK, EGYPT, FRANCE, GERMANY, ICELAND, Italy, JAPAN, JORDAN, Latvia, LUXEMBOURG, MOROCCO, NETHERLANDS, Norway, Oman, PERU, PORTUGAL, Poland, Slovenia, Spain, Switzerland, Sweden, THAILAND, United Kingdom, United Arab Emirates, and VIETNAM.

Code information

Lot Codes: OTW3094¿and OTW3098¿

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0378-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LeMaitre Vascular, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for LeMaitre Vascular, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about LeMaitre Vascular, Inc. →

Source: openFDA record for Z-0378-2016. Report a data issue: hello@regdatalab.com.