Z-0379-2013 Class III Terminated

Recalled by Microgenics Corporation — Fremont, CA

FDA device recall Z-0379-2013 was initiated by Microgenics Corporation on Aug 20, 2012 and is designated Class III. Reason for recall: In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability. The recall status is terminated (terminated Mar 4, 2013). Affected quantity: 2076 total in US , 596 total Foreign.

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Aug 20, 2012
Termination Date
Mar 4, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2076 total in US , 596 total Foreign

Product description

Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .

Reason for recall

In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.

Distribution pattern

Worldwide Distribution--USA (nationwide) and the countries of Germany, Canada and Australia.

Code information

MAS Onmi IMMUNE: Catalog Number/lot number/ expiration date: OIM-101 - OIM14031 - 31 MAR 2014; OIM-202 - OIM14032 - 31 MAR 2014; OIM-303 - OIM14033 - 31 MAR 2014; OIM-SP - OIM1403S - 31 MAR 2014. MAS Omni IMMUNE PRO: OPRO-101 - OPRO14031 - 31 MAR 2014; OPRO-201 - OPRO14032 - 31 MAR 2014' OPRO-303 - OPRO14033 - 31 MAR 2014; OPRO-SP - OPRO1403S - 31 MAR 2014.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0379-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Microgenics Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0379-2013. Report a data issue: hello@regdatalab.com.