Z-0501-2015 Class II Terminated

Recalled by Microgenics Corporation — Fremont, CA

Recall Details

Product Type
Devices
Report Date
December 10, 2014
Initiation Date
October 20, 2014
Termination Date
March 30, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2135 vials.

Product Description

MAS CardioImmune XL Cardiac Marker Control; Model: CAI-XL1; intended for use in the clinical laboratory as an assayed control serum for monitoring assay conditions in specific cardiac marker determinations.

Reason for Recall

MAS CardioImmune XL, level 1 (lot CXL16011), is showing vial-to-vial variation for the analytes Creatine Kinase-MB (CK-MB) and B-Type Natriuretic Peptide-32 (BNP-32).

Distribution Pattern

Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AR, AZ, CA CO, CT, DE, FL GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY., and the countries of Australia, Canada, China, Germany, Guam, Hong Kong and Taiwan.

Code Information

Model No.: CAI-XL1; Lot Number: CXL16011, Exp. Date: 2016-01-31.