Z-1284-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 6, 2022
- Initiation Date
- January 20, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 615 kits
Product Description
QMS Tacrolimus Immunoassay REF 10015573 CALIBRATOR SET QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3
Reason for Recall
Due to QMS Tacrolimus Calibrator recovering between 82 to 94% of their target values, resulting in over-quantitation in tacrolimus patient result values when used with the QMS Tacrolimus Immunoassay.
Distribution Pattern
U.S.: AL, AZ, CA, DC, FL, IA, ID, IL, KY, LA, MA, NE, NJ, NM, OH, OK, PA, SC, TN, TX, and VA O.U.S.: Bangladesh, Brazil, Canada, Germany, India, Malaysia, Mexico, South Korea, and Thailand
Code Information
Model/Part Number: 10015573 UDI Codes: (01) 0884883 01259 2 (17) 220310 (10) 74175462 (01) 0884883 01259 2 (17) 220512 (10) 74267412 Lot Numbers: 74175462 74267412 Calibrators within the set: QMS TACROLIMUS CAL A 10015566-3 QMS TACROLIMUS CAL B 10015567-3 QMS TACROLIMUS CAL C 10015568-3 QMS TACROLIMUS CAL D 10015569-3 QMS TACROLIMUS CAL E 10015570-3 QMS TACROLIMUS CAL F 10015571-3