Z-2884-2016 Class II Terminated

Recalled by Microgenics Corporation — Fremont, CA

FDA device recall Z-2884-2016 was initiated by Microgenics Corporation on Jul 25, 2016 and is designated Class II. Reason for recall: Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL. The recall status is terminated (terminated Nov 21, 2016). Affected quantity: 1708 kits.

Recall details

Product Type
Devices
Report Date
Oct 5, 2016
Initiation Date
Jul 25, 2016
Termination Date
Nov 21, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1708 kits

Product description

DRI Salicylate Serum Tox Assay, Catalog number 0977. Toxicology: The DRI Salicylate Serum Tox Assay is intended for the quantitative determination of salicylate in human serum or plasma.

Reason for recall

Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL.

Distribution pattern

Worldwide Distribution: US (nationwide) to states of: AZ, CA, MO, TX, and PR and countries of: Canada, Germany, Hong Kong, Italy, Malaysia, and Taiwan.

Code information

Lot Number Expiration Date 60704602 DEC 2016; 72035410 JUL 2017; 72208738 FEB 2018; 72409454 OCT 2018.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2884-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Microgenics Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2884-2016. Report a data issue: hello@regdatalab.com.