Z-2884-2016 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 5, 2016
- Initiation Date
- July 25, 2016
- Termination Date
- November 21, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1708 kits
Product Description
DRI Salicylate Serum Tox Assay, Catalog number 0977. Toxicology: The DRI Salicylate Serum Tox Assay is intended for the quantitative determination of salicylate in human serum or plasma.
Reason for Recall
Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL.
Distribution Pattern
Worldwide Distribution: US (nationwide) to states of: AZ, CA, MO, TX, and PR and countries of: Canada, Germany, Hong Kong, Italy, Malaysia, and Taiwan.
Code Information