Z-0379-2018 Class II Ongoing

Recalled by Carl Zeiss Metrology Inc — Maple Grove, MN

FDA device recall Z-0379-2018 was initiated by Carl Zeiss Metrology Inc on Nov 14, 2017 and is designated Class II. Reason for recall: Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the… The recall status is ongoing. Affected quantity: 13 US.

Recall details

Product Type
Devices
Report Date
Feb 7, 2018
Initiation Date
Nov 14, 2017
Voluntary/Mandated
FDA Mandated
Product Quantity
13 US

Product description

Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.

Reason for recall

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Distribution pattern

Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.

Code information

Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0379-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Carl Zeiss Metrology Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Carl Zeiss Metrology Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Carl Zeiss Metrology Inc →

Source: openFDA record for Z-0379-2018. Report a data issue: hello@regdatalab.com.