Z-0379-2018 Class II Ongoing
FDA device recall Z-0379-2018 was initiated by Carl Zeiss Metrology Inc on Nov 14, 2017 and is designated Class II. Reason for recall: Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the… The recall status is ongoing. Affected quantity: 13 US.
Recall details
- Product Type
- Devices
- Report Date
- Feb 7, 2018
- Initiation Date
- Nov 14, 2017
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 13 US
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.