Z-0381-2013 Class II Terminated

Recalled by Stryker Spine — Allendale, NJ

FDA device recall Z-0381-2013 was initiated by Stryker Spine on Aug 9, 2012 and is designated Class II. Reason for recall: Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening. The recall status is terminated (terminated Sep 30, 2015). Affected quantity: 19 units.

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Aug 9, 2012
Termination Date
Sep 30, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 units

Product description

Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Reason for recall

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Distribution pattern

Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

Code information

Class I device, 510(K) exempt Catalog Number: 48287028 Lot Number: 098400

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0381-2013
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0381-2013. Report a data issue: hello@regdatalab.com.