Z-0384-2018 Class II Terminated

Recalled by Valeant Pharmacueticals International — Rochester, NY

FDA device recall Z-0384-2018 was initiated by Valeant Pharmacueticals International on Mar 13, 2017 and is designated Class II. Reason for recall: SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated method, NB-TM-261, In November 2016. The particle size specification set in NB-SPEC-039… The recall status is terminated (terminated Jun 18, 2019). Affected quantity: 418 SPAG-2 and 808 Nebulizers.

Recall details

Product Type
Devices
Report Date
Jan 24, 2018
Initiation Date
Mar 13, 2017
Termination Date
Jun 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
418 SPAG-2 and 808 Nebulizers

Product description

Small Particle Aerosol Generator Model SPAG-2 6000 Series (SPAG-2). The SPAG-2 medical device is a pneumatic flow system indicated for the administration of Virazole (ribavirin) aerosol. Driven by a pressurized gas (air or blended oxygen) and uses a nebulizer to generate a fine aerosol of hydrated Virazole. --- CDRH Classification Name: Nebulizer (Direct Patient Interface)

Reason for recall

SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated method, NB-TM-261, In November 2016. The particle size specification set in NB-SPEC-039 for Dv10, Dv50, & Dv90 are a mean value of no more than 3.0¿m, 5.5 ¿m, and 15¿m respectively. Initial testing of the nebulizers yielded four units that delivered particle sizes that did not meet the specification (NB-SPEC-039). The out of specification (OOS) nebulizers were cleaned per the approved SPAG-2 procedure, SPE-000127228 Rev 2, to remove any product residue and were then re-evaluated. Secondary testing yielded that three of the four nebulizers still did not meet the particle size specification. It was concluded that the 3 nebulizers from SPAG-2 Lot J0571670RA failed to meet (SPE-000127228 Rev 2) did not resolve the high droplet size distribution values observed.

Distribution pattern

US Nationwide & Canada

Code information

Lot Number J0571670RA*/Serial Number 337170600007; Lot Number J0571670R*/Serial Number 337170600007; Lot Number J0571670/Serial Number 337170600007; Lot Number L0197340/Serial Number 340150600003; Lot Number K1900650/Serial Number 340150600003.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0384-2018
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Valeant Pharmacueticals International? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Valeant Pharmacueticals International: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Valeant Pharmacueticals International →

Source: openFDA record for Z-0384-2018. Report a data issue: hello@regdatalab.com.