Z-0388-2027 Class II Ongoing

Recalled by Medtronic Xomed, Inc. — Jacksonville, FL

FDA device recall Z-0388-2027 was initiated by Medtronic Xomed, Inc. on Sep 24, 2026 and is designated Class II. Reason for recall: Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could… The recall status is ongoing. Affected quantity: 2000.

Recall details

Product Type
Devices
Report Date
Nov 18, 2026
Initiation Date
Sep 24, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2000

Product description

BOBBIN VENTILATION TUBE 5PK 1.14 FLPL, REF: 1010201, Lot# 0231048277 (GTIN 20763000281612).

Reason for recall

Ventilation tubes, that provide a passageway for movement of air between the auditory canal and the middle ear, were not subjected to the required sterilization process, which after implantation could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.

Distribution pattern

Worldwide - US Nationwide distribution including in the states of NY, AL, VA, MI, AZ, WA, TX, MA, IL, ME, NM, CA, IN, NC, OH, MO, MN, NH, IA, NV, GA, PA, FL, TN, WV, SC, OK, NE, OR, SD, AK, LA, CO, WI, MD, KS, UT, ID, WY, NJ and the countries of Canada, Australia, New Zealand, Hong Kong.

Code information

UDI-DI: 20763000281612. Lots: 0231048277

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0388-2027
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Xomed, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0388-2027. Report a data issue: hello@regdatalab.com.