Z-0394-2022 Class I Terminated

Recalled by Cardiovascular Systems Inc — Saint Paul, MN

FDA device recall Z-0394-2022 was initiated by Cardiovascular Systems Inc on Nov 22, 2021 and is designated Class I. Reason for recall: Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to… The recall status is terminated (terminated Aug 31, 2023). Affected quantity: 711 units.

Recall details

Product Type
Devices
Report Date
Dec 29, 2021
Initiation Date
Nov 22, 2021
Termination Date
Aug 31, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
711 units

Product description

WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Number) WRN-D6

Reason for recall

Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization.

Distribution pattern

United States- AK, AL, AR, AZ, CA, DC, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV OUS - Netherlands

Code information

Lot Numbers: Lot 67876, UDI 011085000049188217221221100000067876 Lot 70309, UDI 011085000049188217230110100000070309 Lot 68706, UDI 011085000049188217221227100000068706 Lot 69543, UDI 011085000049188217230103100000069543 Lot 75758, UDI 011085000049188217230203100000075758 Lot 77278, UDI 011085000049188217230210100000077278 Lot 79994, UDI 011085000049188217230223100000079994 Lot 81252, UDI 011085000049188217230301100000081252 Lot 81986, UDI 011085000049188217230307100000081986 Lot 83080, UDI 011085000049188217230313100000083080 Lot 81998, UDI 011085000049188217230320100000081998 Lot 84796, UDI 011085000049188217230327100000084796 Lot 87909, UDI 011085000049188217230411100000087909 Lot 89035, UDI 011085000049188217230418100000089035 Lot 86622, UDI 011085000049188217230404100000086622 Lot 90096, UDI 011085000049188217230424100000090096 Lot 93949, UDI 011085000049188217230515100000093949 Lot 91603, UDI 011085000049188217230501100000091603 Lot 94895, UDI 011085000049188217230521100000094895 Lot 91939, UDI 011085000049188217230507100000091939 Lot 101930, UDI 011085000049188217230717100000101930

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0394-2022
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0394-2022. Report a data issue: hello@regdatalab.com.