Z-0395-2017 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0395-2017 was initiated by Roche Diagnostics Operations, Inc. on Oct 4, 2016 and is designated Class II. Reason for recall: Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 4976.

Recall Details

Product Type
Devices
Report Date
Nov 23, 2016
Initiation Date
Oct 4, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4976

Product Description

Tina-Quant Hemoglobin A1c Gen. 2 Hemoglobin A1c test Catalog number 04528123190 / 05401640190

Reason for Recall

Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results. If a patient with known diabetes is affected, an erroneously high HbA1c result might lead to therapeutic consequences, such as therapy escalation (introduction of another oral antidiabetic medication or insulin) or increase of the dosage. This can further lead to an increased risk of hypoglycemia. Considering the unreliable detectability of the issue, relevant medical risk for the patient cannot entirely be excluded.

Distribution Pattern

Nationwide.

Code Information

Not applicable

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0395-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.