Z-0395-2018 Class II Ongoing

Recalled by Measurement Specialties Inc — Andover, MN

FDA device recall Z-0395-2018 was initiated by Measurement Specialties Inc on Mar 13, 2017 and is designated Class II. Reason for recall: The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study. The recall status is ongoing. Affected quantity: 6,076 devices total.

Recall details

Product Type
Devices
Report Date
Jan 31, 2018
Initiation Date
Mar 13, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6,076 devices total

Product description

Measurement Specialties Reusable Temperature Probe Autoclavable. Sold as Esophageal/Rectal (Adult), Esophageal/Rectal (Pediatric), and Skin (Adult) Intended for use by a licensed physician to continuously monitor human esophageal, rectal, or skin temperature. They are designed to be used with any patient monitoring system designed to accept 400 Series temperature probes.

Reason for recall

The reusable temperature probes are provided with instructions for sterilization that have not been proven through a validation study.

Distribution pattern

USA (nationwide) distribution to the states of : AL, CA, KY, IL, KS, MN, NY, OH, TX, VA and WI. Distributed internationally to: Australia, Canada, China, England, Germany, India, Ireland, Israel, Italy, Japan, Malaysia, Netherlands, Singapore, South Africa, South Korea, Taiwan, Vietnam

Code information

Part No. 401AC (Lot No. 14K87246, 15M96863, 13E72645, 12D61386, 13D72302, 16B97593, 15C91286, 13B64377, 14J86547, 12B59416, 13A69334, 15K95972, 15A97196, 15E92377, 15F92378, 13L77336, 12D61023, 14G84704, 14K87499, 14F84193, 13L77335, 16B97594, 15C91419, 13B70167, 15C91448, 15E92375, 15K96046, 16A97197, 12B59529, 16B97798, 15K96125, 16B97595, 15L96777, 12D61829, 13D72644, 15A89998, 13F73449, 12M64134, 15L96745, 15D90215, 15H95035, 13F74403, 13L78489, 13L77336, 12H65961, 14G84878, 12F62179, 14J87014, 12B55182, 14F84444, 14J87081, 13L77336, 13L78489, 14F84400, 12J66632, 16A97380, 12M68130, 15E92376, 14L88737, 14L87904, 12K66633, 13H74994, 13H74995, 12B60056, 12C57285, 14H86266, 16A97592, 14J87164, 12D59168, 14M89765, 12A54328, 15A89854, 14F84634, 14G84624, 15M96829, 15K95896); Part No. 402AC (Lot No. 13E73262, 15E92379, 12E57286, 12L68344, 15E93385, 12H65752, 12H65787, 15F92380, 12J65322, 15F92381, 12K67454, 15A89941, 15A89282, 13H75930, 12F55548, 12K67533, 13H75930, 12J66809, 12B59570, 15E92010, 12A50150, 14E83393); Part No. 409AC (Lot No. 12C56215, 12B56215, 15C90920, 15K28358, 12C60069, 13B70035, 16B97312, 13A69520, 14E83602, 13E73264, 12L68584, 11M57610, 12F59571, 15B90832, 15E92560, 12K63370, 16B97614, 12L54648, 15D92011, 12M63372, 14H85773, 16B97615, 15M97109, 16B97805, 12M69229, 15A89939, 15A90196, 12F60692, 14L88520, 14L88519, 13A70145, 15E92382, 15A90466, 13E73450, 15F92383, 13D72925)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0395-2018
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Measurement Specialties Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Measurement Specialties Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Measurement Specialties Inc →

Source: openFDA record for Z-0395-2018. Report a data issue: hello@regdatalab.com.