Z-0396-2017 Class II Terminated

Recalled by Roche Diagnostics Operations, Inc. — Indianapolis, IN

FDA device recall Z-0396-2017 was initiated by Roche Diagnostics Operations, Inc. on Oct 4, 2016 and is designated Class II. Reason for recall: .Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred… The recall status is terminated (terminated Jun 5, 2017). Affected quantity: 4976.

Recall Details

Product Type
Devices
Report Date
Nov 23, 2016
Initiation Date
Oct 4, 2016
Termination Date
Jun 5, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4976

Product Description

COBAS INTEGRA 800 Tina-Quant Hemoglobin A1cDX Gen.2 Hemoglobin A1cDX test Catalog number 04528123160

Reason for Recall

.Roche Diagnostics has confirmed elevated QC and patient sample recovery for individual Tina-quant¿ Hemoglobin A1c Gen.2 and Tina-quant Hemoglobin A1cDx Gen.2 cassettes/bottles. This issue occurred over multiple lots. Single cassettes/bottles of the reagents are affected. The issue can lead to erroneously high HbA1c results.

Distribution Pattern

Nationwide.

Code Information

Not applicable

Source

Dataset
openFDA device recall enforcement reports