Z-0401-2017 Class III Terminated
FDA device recall Z-0401-2017 was initiated by Siemens Healthcare Diagnostics, Inc. on Aug 25, 2016 and is designated Class III. Reason for recall: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU). The recall status is terminated (terminated Dec 1, 2017). Affected quantity: 30,247 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 23, 2016
- Initiation Date
- Aug 25, 2016
- Termination Date
- Dec 1, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 30,247 units
Product description
Dimension Vista IRON Flex reagent cartridges
Reason for recall
Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).
Distribution pattern
Distributed Nationwide
Code information
Device Listing # D011391 Lot # 15244AB, 15301BA,15345BB,16026BA,16075BB,16117BD,16166BA,16201BD Exp Date 08-31-2016 - 07/19-2016.