Z-0402-2014 Class II Terminated

Recalled by Advanced Sterilization Products — Irvine, CA

FDA device recall Z-0402-2014 was initiated by Advanced Sterilization Products on Nov 11, 2013 and is designated Class II. Reason for recall: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the… The recall status is terminated (terminated Jan 22, 2014). Affected quantity: Total 4,381 units.

Recall details

Product Type
Devices
Report Date
Dec 4, 2013
Initiation Date
Nov 11, 2013
Termination Date
Jan 22, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total 4,381 units

Product description

STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Reason for recall

Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

Distribution pattern

Nationwide Distribution

Code information

Please refer to consignee list

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0402-2014
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Advanced Sterilization Products? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0402-2014. Report a data issue: hello@regdatalab.com.