Z-0402-2017 Class III Terminated
FDA device recall Z-0402-2017 was initiated by Siemens Healthcare Diagnostics, Inc. on Aug 25, 2016 and is designated Class III. Reason for recall: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU). The recall status is terminated (terminated Dec 1, 2017). Affected quantity: 38,230 cartridges.
Recall details
- Product Type
- Devices
- Report Date
- Nov 23, 2016
- Initiation Date
- Aug 25, 2016
- Termination Date
- Dec 1, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 38,230 cartridges
Product description
Dimension IRON Flex reagent cartridges
Reason for recall
Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).
Distribution pattern
Distributed Nationwide
Code information
Device Listing # D011391 Lot # DB6267, FD6309, FB6310,BA7006,BB7006,GB7060,BA7116,EC7200 Exp Date: 09/23/2016-0-7/19/2017
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0402-2017
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026