Z-0402-2017 Class III Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

FDA device recall Z-0402-2017 was initiated by Siemens Healthcare Diagnostics, Inc. on Aug 25, 2016 and is designated Class III. Reason for recall: Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU). The recall status is terminated (terminated Dec 1, 2017). Affected quantity: 38,230 cartridges.

Recall details

Product Type
Devices
Report Date
Nov 23, 2016
Initiation Date
Aug 25, 2016
Termination Date
Dec 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
38,230 cartridges

Product description

Dimension IRON Flex reagent cartridges

Reason for recall

Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instructions for Use (IFU).

Distribution pattern

Distributed Nationwide

Code information

Device Listing # D011391 Lot # DB6267, FD6309, FB6310,BA7006,BB7006,GB7060,BA7116,EC7200 Exp Date: 09/23/2016-0-7/19/2017

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0402-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0402-2017. Report a data issue: hello@regdatalab.com.