Z-0404-2020 Class II Terminated

Recalled by Galil Medical, Inc. — Saint Paul, MN

FDA device recall Z-0404-2020 was initiated by Galil Medical, Inc. on Oct 25, 2019 and is designated Class II. Reason for recall: Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance. The recall status is terminated (terminated Sep 16, 2020). Affected quantity: 49 systems.

Recall details

Product Type
Devices
Report Date
Nov 20, 2019
Initiation Date
Oct 25, 2019
Termination Date
Sep 16, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
49 systems

Product description

Visual-ICE Cryoablation System, FPRCH6000. The cryoablation system is a mobile console system intended for cryoablative tissue destruction using a minimally invasive procedure. The system is computer-controlled with a touch screen user interface that allows the user to control and monitor the procedure. The therapy delivered by the system is based on the Joule-Thomson effect displayed by compressed gases. The Visual-ICE System uses high-pressure argon gas that circulates through closed-tip cryoablation needles to induce tissue freezing. Active tissue thawing is achieved by circulating helium gas through the needles or, alternatively , by the use of Galil Medical I-Thaw technology in which a heating element inside the cryoablation needle can be energized to cause thawing.

Reason for recall

Visual-ICE utilizing the IceSeed 1.5 and IceRod 1.5 PLUS cryoablation needles could have issues with the needle clogging and/or inadequate freezing performance.

Distribution pattern

AK, AL, AR, AZ, CA, FL, GA, HI, KY, MA, MI, MN, NC, ND, NY, OH, PA, SC, TN, UT, VA, Puerto Rico OUS to include: Argentina, Canada, Taiwan, United Kingdom

Code information

System Serial Number / PM Kit #: VL0031 / PM0506M, VL0036 / PM0546M, VL0067 / PM0515M, VL0069 / PM0565M, VL0075 / PM0532M, VL0079 / PM0539M, VL0082 / PM0491M, VL0085 / PM0486M, VL0086 / PM0499M, VL0088 / PM0489M, VL0090 / PM0513M, VL0092 / PM0548M, VL0093 / PM0544M, VL0094 / PM0520M, VL0100 / PM0516M, VL0102 / PM0529M, VL0105 / PM0560M, VL0112 / PM0542M, VL0115 / PM0525M, VL0136 / PM0561M, VL0148 / PM0514M, VL0153 / PM0531M, VL0171 / PM0564M, VL0175 / PM0559M, VL0181 / PM0512M, VL0186 / PM0533M, VL0191 / PM0492M, VL0220 / PM0545M, VL0230 / PM0528M, VL0238 / PM0500M, VL0249 / PM0498M, VL0271 / PM0497M, VL0275 / PM0543M, VL0280 / PM0549M, VL0283 / PM0488M, VL0285 / PM0566M, VL0287 / PM0562M, VL0291 / PM0524M, VL0300 / PM0550M, VL0301 / PM0541M, VL0303 / PM0530M, VL0304 / PM0563M, VL0308 / PM0523M, VL0309 / PM0507M, VL0310 / PM0490M, VL0311 / PM0555M, VL0312 / PM0526M, VL0321 / PM0552M, VL0362 / PM0527M.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0404-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Galil Medical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Galil Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Galil Medical, Inc. →

Source: openFDA record for Z-0404-2020. Report a data issue: hello@regdatalab.com.