Z-0410-2025 Class II Ongoing

Recalled by AVID Medical, Inc. — Toano, VA

FDA device recall Z-0410-2025 was initiated by AVID Medical, Inc. on Sep 20, 2024 and is designated Class II. Reason for recall: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing… The recall status is ongoing. Affected quantity: 22 eaches.

Recall details

Product Type
Devices
Report Date
Nov 20, 2024
Initiation Date
Sep 20, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 eaches

Product description

Halyard KNEE ARTHROSCOPY PACK - Medical convenience kits Model Number: MAMC204-16

Reason for recall

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Distribution pattern

US Nationwide distribution.

Code information

GTIN: 10809160360767 Lot Number: 1608931

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0410-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0410-2025. Report a data issue: hello@regdatalab.com.