Z-0415-2025 Class II Ongoing

Recalled by AVID Medical, Inc. — Toano, VA

FDA device recall Z-0415-2025 was initiated by AVID Medical, Inc. on Sep 20, 2024 and is designated Class II. Reason for recall: Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing… The recall status is ongoing. Affected quantity: 44 eaches.

Recall Details

Product Type
Devices
Report Date
Nov 20, 2024
Initiation Date
Sep 20, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
44 eaches

Product Description

Halyard ORTHO PACK -Medical convenience kits Model Number: MMOK006-04

Reason for Recall

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Distribution Pattern

US Nationwide distribution.

Code Information

GTIN: 10809160404201 Lot Number: 1597105,1605626

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0415-2025
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.