Z-0417-2013 Class II Terminated

Recalled by SpineFrontier, Inc. — Beverly, MA

FDA device recall Z-0417-2013 was initiated by SpineFrontier, Inc. on Aug 22, 2012 and is designated Class II. Reason for recall: When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating… The recall status is terminated (terminated Jun 11, 2015). Affected quantity: 19 units.

Recall details

Product Type
Devices
Report Date
Nov 28, 2012
Initiation Date
Aug 22, 2012
Termination Date
Jun 11, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 units

Product description

S-Lift Implant Guides (also known as S-LIF) Compressor Part Number: 11-50223 Product Usage: Surgical Guides allow a surgeon to place implants into tight spaces, such as a vertebral disc space, without having the implant contact the surrounding bone or tissue. In the case of S-LIFT Implant Guides, they are used to place vertebral bodies into the disc space without the body contacting the superior or inferior vertebrae.

Reason for recall

When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating the potential for injury to the patient. The instructions for Use for the S-Lift Implant Guides will be amended to add the following warning: "Care is to be taken to not over insert the Implant Guides

Distribution pattern

US Nationwide including the states of: MA and TX.

Code information

Lot Numbers: 012064-001

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0417-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0417-2013. Report a data issue: hello@regdatalab.com.