Z-0419-2022 Class II Ongoing

Recalled by Intuitive Surgical, Inc. — Sunnyvale, CA

Recall Details

Product Type
Devices
Report Date
January 5, 2022
Initiation Date
September 22, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
46

Product Description

da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A

Reason for Recall

Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.

Distribution Pattern

US Nationwide distribution in the states of Arizona, California, Connecticut, Florida, Illinois, Kansas, Kentucky, Massachusetts, Minnesota, Missouri, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington, and Wisconsin.

Code Information

All da Vinci SP Surgical systems with software version OS4 v4.0.0 (P4_B327)